Aquascat AOBS Super-Turbidity sensor

The AQUAscat ® AOBS is a simple, low-cost Super-Turbidity sensor designed to measure suspended sediment concentration (SSC). The Super-Turbidity technology involves pairing an AQUAscat ® ABS instrument with a turbidity sensor using a weight factor, which results in a single, combined output from the two sensors. Once paired, the AQUAscat ® AOBS retains near-constant calibration for SSC over a wide grain-size range. The AQUAscat ® AOBS Super-Turbidity sensor is supplied as an integrated and paired turbidity and acoustic sensor with a variety of cabling and data logger options

The TECPRIME AQUAscat® AOBS is an integrated acoustic and optical Super-Turbidity sensor designed to measure suspended sediment concentration across a broad range of particle sizes.

Formerly known as the Sequoia Scientific LISST-AOBS, the instrument is now manufactured, sold and supported by TECPRIME Technology under the AQUAscat AOBS name. Its measurement technology is licensed from Sequoia Scientific.

The AQUAscat AOBS combines an 8 MHz acoustic backscatter sensor with an 850 nm optical turbidity sensor. These two measurement methods respond differently to changes in sediment grain size. Optical turbidity is particularly responsive to fine particles, while high-frequency acoustic backscatter provides a comparatively stable response to coarse silt and sand.

By applying a predetermined weighting factor to the acoustic and optical measurements, the AQUAscat AOBS produces a combined suspended sediment concentration output with substantially less grain-size sensitivity than either technology used independently. TECPRIME specifies near-constant calibration within a factor of two for grain sizes from approximately 1-500 µm.

The complete system includes an AQUAscat ABS acoustic sensor, a Turner Designs Turbidity Plus optical sensor, an integrated mounting arrangement and a Y-cable providing power and digital communications to both sensors.

Imbros can assist with selecting the appropriate AQUAscat AOBS cabling, datalogger, telemetry, mounting and calibration arrangement for fixed monitoring, profiling and research applications.

Why Choose the AQUAscat AOBS?

  • – Combined acoustic and optical measurement: integrates two complementary technologies into one suspended sediment monitoring system.
  • – Broad particle-size response: designed to maintain calibration within a factor of two over grain sizes from approximately 1–500 µm.
  • – Direct suspended sediment output: the weighted Super-Turbidity result is reported as suspended sediment concentration rather than turbidity alone.
  • – Measures fine and coarse sediment: the optical sensor responds strongly to fine material, while the acoustic sensor provides improved sensitivity to coarse silt and sand.
  • – Wide measurement range: the acoustic element has a published suspended sediment range of approximately 1–30,000 mg/L, depending on sediment characteristics.
  • – Integrated system: supplied as a paired AQUAscat ABS and Turner Designs Turbidity Plus sensor package.
  • – Mechanical optical wiper: the Turbidity Plus sensor includes an integrated wiper to help maintain the optical measurement surface.
  • – Fouling-tolerant acoustic measurement: the acoustic sample volume is positioned away from the transducer face and does not rely on an optical window.
  • – Digital integration: SDI-12 and RS-232 outputs support connection to environmental dataloggers and monitoring systems.
  • – Suitable for variable sediment: particularly useful where the particle-size distribution changes during storms, floods, dredging or seasonal flow events.

AQUAscat AOBS at a Glance

AttributePublished Specification
Instrument typeIntegrated acoustic and optical Super-Turbidity sensor
Primary measurementSuspended sediment concentration
Additional measurementOptical turbidity output
Acoustic sensorAQUAscat ABS operating at 8 MHz
Optical sensorTurner Designs Turbidity Plus using an 850 nm optical source
Optical measurement methodISO 7027 turbidity technique
Published grain-size responseNear-constant calibration within a factor of two from approximately 1–500 µm
Acoustic SSC rangeApproximately 1–30,000 mg/L, particle dependent
Optical turbidity rangeUp to 3,000 NTU following user calibration
Acoustic sample volumeApproximately 10 mm diameter × 15 mm long
Acoustic sample locationApproximately 55 mm in front of the acoustic transducer
OutputsSDI-12 and RS-232
Depth rating100 m
Operating temperature0–30 °C; not recommended for use in water bodies that freeze
Dimensions57.2 × 101.6 × 336.5 mm
Weight in airApproximately 0.7 kg

The optical Turbidity Plus sensor is not supplied factory-calibrated in NTU. A user calibration is required where a traceable turbidity result is needed. Quantitative suspended sediment performance also depends on sediment size, density, shape and composition.

Typical Applications

ApplicationHow the AQUAscat AOBS Can Be Used
River and estuary monitoringContinuous measurement of suspended sediment during changing flows, tides, floods and catchment-runoff events where both fine and coarse material may be present.
Dredge plume monitoringMonitoring sediment released by dredging, reclamation, excavation and disposal activities across a changing particle-size distribution.
Hydropower sediment monitoringMeasuring fine and abrasive coarse sediment entering reservoirs, channels, intakes or turbine systems.
Coastal erosion studiesObserving the suspension and transport of sediment generated by waves, currents, erosion and shoreline-management works.
Environmental impact assessmentsCollecting continuous sediment data before, during and after construction, dredging or other potentially disturbing activities.
Flood and storm-event monitoringCapturing rapid changes in sediment concentration and grain-size composition that may not be represented by periodic grab samples.
Reservoir and catchment studiesQuantifying sediment inflows, resuspension and transport through reservoirs and impoundments.
Sediment transport researchInvestigating the relationship between suspended concentration, grain size, flow conditions and sediment movement.
Total suspended solids studiesDeveloping site-specific relationships between continuous AOBS measurements and laboratory filtration or gravimetric results.
Scientific researchProviding simultaneous acoustic, optical and combined measurements for comparison with water samples and other sediment instruments.

What Is Super-Turbidity?

Conventional turbidity sensors measure light scattered by particles in the water. Their response is normally strongest for fine particles, while the response per unit mass can decrease substantially as particles become larger.

An acoustic backscatter sensor behaves differently. The 8 MHz AQUAscat ABS has a comparatively consistent response to particles from approximately 30–400 µm, but its sensitivity decreases as sediment becomes finer than this range.

The opposing grain-size responses of the two technologies make it possible to combine them:

  • – The optical sensor provides strong sensitivity to fine clay and silt;
  • – The acoustic sensor responds effectively to coarse silt and sand; and
  • – A weighting factor is applied to combine both measurements into one suspended sediment result.

The resulting Super-Turbidity measurement is substantially less sensitive to changes in grain size than either the optical or acoustic measurement alone.

Why this matters: a river monitoring site may contain fine suspended clay during normal flow and substantially coarser material during a flood. A single optical or acoustic sensor may change sensitivity as the sediment changes. The AQUAscat AOBS is designed to reduce this source of measurement uncertainty.

How the Two Sensors Work Together

The AQUAscat AOBS incorporates two physically separate but integrated sensors.

8 MHz Acoustic Backscatter Sensor

The AQUAscat ABS transmits an ultrasonic pulse into the water and measures the sound scattered back by particles in a small sample volume approximately 55 mm in front of the transducer.

The acoustic sensor is particularly useful for measuring coarse silt and sand. It also measures acoustic attenuation so that the backscatter result can be corrected when higher sediment concentrations weaken the transmitted and returned signal.

850 nm Optical Turbidity Sensor

The Turner Designs Turbidity Plus sensor measures scattered near-infrared light using an optical technique based on ISO 7027. It is particularly responsive to fine suspended particles.

The optical sensor incorporates a mechanical wiper to reduce the accumulation of biological growth and sediment on the optical surface.

Weighted Combined Output

The sensors are paired using a weighting factor derived from their relative response to a selected reference particle size. The weighted optical measurement is added to the acoustic measurement to create the combined Super-Turbidity suspended sediment output.

The optical units used during the pairing process do not need to be NTU. The weighting method uses the relative optical response, and the resulting combined measurement is reported as suspended sediment concentration.

AQUAscat AOBS Measurement Outputs

OutputPurposeImportant Consideration
Acoustic measurementProvides the AQUAscat ABS response to suspended particles, with strongest application across coarse silt and sand.Response becomes more grain-size dependent for particles below approximately 30 µm.
Optical turbidity measurementProvides a sensitive optical response to fine suspended material.The Turbidity Plus sensor is not factory-calibrated in NTU and requires user calibration where NTU output is required.
Combined Super-Turbidity SSCCombines the acoustic and weighted optical responses into one suspended sediment concentration result.Designed to maintain calibration within a factor of two across approximately 1–500 µm.

Measurement, Calibration and Installation

Particle-Size Range and Measurement Performance

The principal advantage of the AQUAscat AOBS is its ability to accommodate a wider particle-size distribution than either acoustic backscatter or optical turbidity used independently.

TECPRIME specifies near-constant calibration within a factor of two for grain sizes from approximately 1–500 µm. This range includes:

  • – Clay-sized particles
  • – Fine and coarse silt
  • – Fine sand
  • – Medium sand near the upper end of the published range

The factor-of-two specification should not be interpreted as laboratory-grade gravimetric accuracy under every condition. Sediment density, shape, mineral composition, organic content, flocculation and mixed particle-size distributions can all affect acoustic and optical response.

Where the project requires defensible quantitative concentration data, field measurements should be supported by representative water samples and independent laboratory analysis.

Suspended Sediment and Turbidity Calibration

The AQUAscat AOBS is supplied as a paired acoustic and optical system. The weighting factor used to combine the two technologies is established as part of the Super-Turbidity configuration.

However, users should distinguish between three different forms of calibration:

Sensor Pairing

Pairing establishes the weighting factor used to combine the AQUAscat ABS and Turbidity Plus outputs. This is fundamental to producing the Super-Turbidity result.

Suspended Sediment Validation

Representative water samples can be collected and analysed using gravimetric or filtration methods to assess how closely the instrument output matches the actual sediment at the monitoring site.

NTU Calibration

The Turner Designs Turbidity Plus sensor is not supplied factory-calibrated in NTU. Where an independent NTU result is required, the optical sensor must be calibrated using a suitable turbidity standard and an appropriate procedure.

Important: NTU calibration of the optical sensor is not necessary for the Super-Turbidity pairing method. It is only required when the user also needs the optical result reported as calibrated turbidity.

Fixed Monitoring Installations

The AQUAscat AOBS is well suited to fixed monitoring stations where sediment characteristics may change over time.

Potential mounting locations include:

  • – River-bank or bridge-mounted monitoring frames
  • – Hydropower channels and intakes
  • – Reservoir monitoring structures
  • – Dredging compliance stations
  • – Coastal and estuarine moorings
  • – Construction monitoring platforms
  • – Submerged instrument cages

The system should be installed so that both sensors measure representative water without interference from the mounting structure.

Installation design should consider:

  • – The location of the acoustic sample volume
  • – Clearance around the optical sensor
  • – The dominant direction of flow
  • – Sediment deposition and burial risk
  • – Debris and vegetation
  • – Changing water levels
  • – Access for optical-wiper inspection
  • – Cable protection and strain relief
  • – Datalogger and telemetry power requirements

An optional mounting frame is available separately from TECPRIME.

Dredging and Construction Monitoring

Dredging and marine construction can produce suspended material ranging from fine clay and silt through to coarser sand. The proportion of each fraction can change as equipment moves between sediment layers or as currents transport the plume.

This variability can make a single turbidity-to-concentration relationship unreliable. The AQUAscat AOBS addresses this by combining:

  • – Optical sensitivity to fine plume material; and
  • – Acoustic sensitivity to coarse suspended sediment.

For compliance and environmental-impact monitoring, the complete system may include:

  • – One or more AQUAscat AOBS instruments
  • – A SDI-12-compatible datalogger
  • – Cellular, radio or satellite telemetry
  • – Solar or mains power
  • – Water-level and current sensors
  • – Automated alarms or reporting
  • – Periodic laboratory water samples

Project consideration: regulatory limits may be expressed as turbidity, total suspended solids or suspended sediment concentration. The required reporting unit should be confirmed before the monitoring system and calibration programme are specified.

Datalogger and Telemetry Integration

The AQUAscat AOBS provides SDI-12 and RS-232 communications for connection to environmental dataloggers and integrated monitoring systems.

SDI-12

SDI-12 is commonly used in environmental monitoring because multiple low-power sensors can share one compatible datalogger input. It is particularly suitable for remote stations incorporating several water-quality and meteorological instruments.

RS-232

RS-232 supports direct serial communication with compatible computers, controllers and datalogging equipment.

Integrated Y-Cable

The supplied Y-cable distributes power and communications to the AQUAscat ABS and Turbidity Plus sensors, allowing the paired system to operate as an integrated instrument package.

The complete monitoring arrangement may also include:

  • – Submersible or shore-based datalogger
  • – Cellular modem
  • – Radio telemetry
  • – Satellite telemetry
  • – Solar charging system
  • – External battery
  • – Waterproof field enclosure
  • – Real-time web portal or alarm system

Need help with integration? Imbros can review the intended datalogger, telemetry method, supply voltage and deployment depth to assist with specifying the appropriate cable and connection arrangement.

Biofouling and Maintenance

The two sensing technologies require different maintenance considerations.

Acoustic Sensor

The AQUAscat ABS does not require an optical measurement window. Its acoustic sample volume is located approximately 55 mm in front of the transducer, making it comparatively tolerant of moderate surface fouling.

Marine growth, debris or sediment deposits can still obstruct the acoustic path and should be removed during scheduled maintenance.

Optical Sensor

The Turbidity Plus includes a mechanical wiper that cleans the optical surface. Wiping frequency should be selected according to deployment conditions and expected fouling pressure.

The wiper reduces maintenance requirements but does not eliminate them. The optical surface, wiper blade and sensor body should be inspected periodically.

Suggested Maintenance Checks

  • – Inspect and clean both sensor bodies
  • – Confirm that the acoustic measurement path is unobstructed
  • – Inspect the optical window and wiper blade
  • – Check the integrated mounting bracket
  • – Inspect cables, connectors and strain relief
  • – Compare readings with representative water samples
  • – Review signal stability and baseline changes
  • – Confirm that sediment has not buried or blocked either sensor

Technical Specifications

Technical ParameterPublished Specification
SensorsCombined 850 nm optical turbidity sensor and 8 MHz acoustic backscatter sensor
Acoustic sensorTECPRIME AQUAscat ABS
Optical sensorTurner Designs Turbidity Plus
Optical wavelength850 nm
Optical techniqueISO 7027 turbidity technique
Optical cleaningIntegrated mechanical wiper
Acoustic frequency8 MHz
Acoustic transducer10 mm diameter ceramic transducer
Acoustic sample volume10 mm diameter × 15 mm long
Sample-volume locationApproximately 55 mm in front of the acoustic sensor
Published grain-size performanceNear-constant calibration within a factor of two from approximately 1–500 µm
Acoustic SSC rangeApproximately 1–30,000 mg/L, sediment dependent
Optical range0–3,000 NTU following calibration
Digital outputsSDI-12 and RS-232

Mechanical and Electrical Specifications

ParameterPublished Specification
Dimensions57.2 mm high × 101.6 mm wide × 336.5 mm long
Weight in airApproximately 0.7 kg
Supply voltage9–15 VDC, 12 VDC nominal
Current drain75 mA at 12 VDC
Depth rating100 m
Operating temperature0–30 °C
Freezing-water suitabilityNot recommended for use in water bodies that freeze
Current connectorSubConn MCBH8M
Legacy connectorImpulse MCBH-8-MP on AQUAscat ABS serial numbers 6233 and lower
Mounting frameOptional; available separately

Connector and cable requirements should be checked against the instrument serial number. Legacy LISST-AOBS cables and documentation should not be assumed to match every current AQUAscat AOBS configuration.

AQUAscat ABS or AQUAscat AOBS?

ConsiderationAQUAscat ABSAQUAscat AOBS
Measurement technology8 MHz acoustic backscatterCombined 8 MHz acoustic backscatter and 850 nm optical turbidity
Best-suited grain sizesApplications dominated by particles from approximately 30–400 µmApplications with grain sizes varying broadly from approximately 1–500 µm
Fine-particle sensitivityReduced below approximately 30 µmImproved through the integrated optical sensor
Coarse-particle sensitivityStrong response across coarse silt and sandRetained through the integrated acoustic sensor
Turbidity measurementNo optical turbidity measurementOptical turbidity available following user calibration
Optical maintenanceNo optical window or wiperOptical sensor with integrated mechanical wiper
System complexityCompact single-sensor packagePaired two-sensor package with integrated cabling
Recommended whenSediment is predominantly coarse or low-maintenance acoustic monitoring is preferredSediment grain size is broad, variable or poorly characterised

Selection guidance: the AQUAscat AOBS is generally the stronger option where both fine and coarse particles contribute materially to suspended sediment concentration. The simpler AQUAscat ABS may remain preferable where sediment is predominantly coarse and the optical measurement is unnecessary.

More information from Tecprime