GloveOn Celeste is a popular choice from our nitrile exam glove range. With high tensile strength and resistance to chemicals, GloveOn Celeste’s superior tactility commends it to delicate procedures.
| GLOVE PROPERTIES | MEASUREMENT (MM) |
| Length | ≥230 |
| Thickness at Palm (centre of Palm) | 0.07 ± 0.02 |
| Thickness at Finger (13mm ± 3mm from tip) | 0.09 ± 0.02 |
| PHYSICAL PROPERTIES | BEFORE AGEING | AFTER AGEING |
| Tensile Strength (MPa) | ≥18 | ≥16 |
| Elongation (%) | ≥500 | ≥400 |
| INSPECTION LEVELS AND AQL | INSPECTION LEVEL | AQL |
| Watertightness | G1 | 1.5 |
| Physical Dimensions | S2 | 4 |
| Tensile Strength | S2 | 4 |
| Visual Inspection (Major) | S4 | 2.5 |
| Visual Inspection (Minor) | S4 | 4 |
| Particulate Residue | N=5 | ≤2mg/glove |
| PACKAGING | |
| Configuration | 100 gloves per box for XS to L |
| 90 gloves per box for XL | |
| 10 boxes per carton |
| REGULATORY COMPLIANCE | |
| Regulations | ARTG 164563 |
| FDA 510(k) | |
| REACH | |
| EU 2016/425 | |
| EU 10/2011 | |
| EC 1935/2004 | |
| MDD 93/42/EEC | |
| ROHS Directive 2002/95/EC |
| INTERNATIONAL STANDARDS | |
| Standards | ASTM D6319 |
| ASTM D6124 | |
| ASTM D5151 | |
| ASTM F1671 | |
| ASTM D6978 | |
| EN 455 part 1, 2, 3 & 4 | |
| EN 1186 | |
| EN 13130 | |
| CEN/TS 14234 | |
| EN 420 | |
| EN ISO 374-1 (Type B) | |
| EN 16523-1 | |
| EN 374 part 2, 4 & 5 | |
| ISO 10993 part 5 & 10 | |
| HACCP Certified | |
| MANUFACTURING ACCREDITATIONS | |
| Accreditations | ISO 9001 |
| ISO 13485 | |
| EN ISO 13485 |
| CHEMOTHERAPY DRUGS AND CONCENTRATION | MINIMUM BREAKTHROUGH DETECTION TIME (MINUTES) |
| (TESTED FOR RESISTANCE TO PERMEATION BY CHEMOTHERAPY DRUGS AS PER ASTM D6978-05, TEST REPORT PN 134889A) | |
| Carmustine (BCNU), 3.3mg/ml (3,300ppm) | 16.2 |
| Cisplatin, 1.0mg/ml (1,000ppm) | >240 |
| Cyclophosphamide (Cytoxan), 20.0mg/ml (20,000ppm) | >240 |
| Dacarbazine (DTIC), 10.0mg/ml (10,000ppm) | >240 |
| Doxorubicin Hydrochloride, 2.0mg/ml (2,000ppm) | >240 |
| Etoposide (Toposar), 20.0mg/ml (20,000ppm) | >240 |
| Fluorouracil, 50.0mg/ml (50,000ppm) | >240 |
| Methotrexate, 25.0mg/ml (25,000ppm) | >240 |
| Mitomycin C, 0.5mg/ml (500ppm) | >240 |
| Paclitaxel (Taxol), 6.0mg/ml (6,000ppm) | >240 |
| Thiotepa, 10.0mg/ml (10,000ppm) | 28.4 |
| Vincristine Sulfate, 1.0mg/ml (1,000ppm) | >240 |
| WARNING: Carmustine and Thiotepa, at the tested concentration, degraded a Celeste nitrile glove at 16.2 minutes and 28.4 minutes respectively. The safe use of gloves in chemotherapy treatment is solely the decision of clinicians authorised to make such a decision. | |
| CATALOGUE NUMBER | SIZE |
|---|---|
| NTR58XS | X-Small |
| NTR58SS | Small |
| NTR58MM | Medium |
| NTR58LL | Large |
| NTR58XL | X-Large |